Instrument:
HP1090 with Diode Array
Column:
Targa C18 5µm 100x2.1mm
P/N: H TS-1021-C185
Mobile Phase:
3% Water in MeOH
Flowrate:
0.2mL/min
Sample:
Vitamin Mix, 2µL injection
Column Temp:
40° C
Detector:
DAD, Optimized Wavelength for each compound
Wavelength Optimization:
Vitamin K3 - 250nm, Vitamin A - 325nm,
Vitamin D3 - 265nm, Vitamin E - 290nm, Vitamin K1 - 245nm
Higgins Analytical is grateful to David Isom of Analytical Sales and Services for sharing these data.
© 1998 Higgins Analytical, Inc. (22Sept98)